FDA Adverse Event Malfunction Summary report: N

TPRLC 133 FP TYPE1 PPS SO 4.0

MDR report key: 17995834 · Received October 24, 2023

Report

Report Number
0001825034-2023-02476
Event Type
Malfunction
Date Received
October 24, 2023
Date of Event
October 5, 2023
Report Date
January 29, 2024
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LPH
UDI-DI
00880304512313
PMA / PMN Number
K120030
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. UPDATED: D9, G3, G6, H2, H3, H6 VISUAL EVALUATION OF THE RETURNED PRODUCT/PROVIDED PHOTOS IDENTIFIED DAMAGE TO THE STERILE PACKAGING (BLISTER). STERILITY HAS NOT BEEN COMPROMISED. THE REPORTED EVENT HAS BEEN CONFIRMED BY EVALUATION OF THE RETURNED PRODUCT AND PROVIDED PHOTOS. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING RELATED TO THE REPORTED EVENT. THE REPORTED EVENT DID NOT OCCUR IN AN OPERATING ROOM OR AS PART OF A MEDICAL PROCEDURE; MEDICAL RECORDS ARE NOT AVAILABLE FOR REVIEW. THE CONDITION OF THE DEVICE WHEN IT LEFT ZIMMER BIOMET IS CONSIDERED CONFORMING TO SPECIFICATION. THE ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO TRANSIT DAMAGE AND A PACKAGING DESIGN ISSUE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

(B)(4). D10: 51-145160 TPRLC XR MP T1 PPS 16X117MM 6541707, 51-105130 TPRLC XR T1 PPS 13X146MM 6883665. G2: FOREIGN: JAPAN. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034 -2023 -02477, 0001825034-2023-02478. CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

Description of Event or Problem · 0

IT WAS REPORTED DURING CIRCULATION THAT THE STERILE PACKAGE OF THE PRODUCTS WAS DAMAGED. IT WAS REPORTED THAT NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
744252 TPRLC 133 FP TYPE1 PPS SO 4.0 PROTHESIS, HIPS LPH ZIMMER BIOMET, INC. N/A 6644495 00880304512313

Patients

Seq Age Sex Outcome Treatment
1 Unknown SEE H10 NARRATIVE.