FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 17993291 · Received October 23, 2023

Report

Report Number
1221359-2023-01570
Event Type
Malfunction
Date Received
October 23, 2023
Date of Event
October 1, 2023
Report Date
October 23, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

D4-UDI:(B)(4). TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 225141 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160/ LOT: 225141, TEST BASE PART NUMBER 195-430WJR/ LOT: 222477. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 225141 SHOWED THAT THE COMPLAINT RATE IS 0.000325%. ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2023 ON A NASAL KITTED SWAB. REPEAT TESTING WAS NOT PERFORMED. INITIAL TESTING WAS PERFORMED ON (B)(6) 2023 VIA AN UNKNOWN BRAND OF RAPID ANTIGEN TEST WHICH GENERATED A NEGATIVE RESULT. ADDITIONAL TESTING WAS PERFORMED ON (B)(6) 2023 VIA AN IHEALTH RAPID ANTIGEN TEST WHICH GENERATED A POSITIVE RESULT. THE CONSUMER STATED THEY EXPERIENCED SYMPTOMS SUCH AS CHILLS AND HIGH FEVER THAT STARTED (B)(6) 2023 AND LASTED FOR TWO (2) DAYS. THE CONSUMER CONFIRMED THERE WAS NO HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CONSUMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
788458 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 225141 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 60 YR Female