FDA Adverse Event Injury Summary report: N

STELLANT DUAL CT INJECTOR

MDR report key: 17986637 · Received October 23, 2023

Report

Report Number
2520313-2023-00026
Event Type
Injury
Date Received
October 23, 2023
Date of Event
September 25, 2023
Report Date
September 23, 2024
Manufacturer
BAYER MEDICAL CARE INC.
Product Code
DXT
UDI-DI
00616258020860
PMA / PMN Number
K082905
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A SYSTEM SERVICE CHECK OF THE MEDRAD® STELLANT CT INJECTION SYSTEM, SERIAL NUMBER (B)(6), WAS DECLINED BY THE CUSTOMER. THE CUSTOMER DISCARDED THE SUSPECT DISPOSABLES AT THE TIME OF THE INCIDENT; HOWEVER, THEY WERE ABLE TO PROVIDE A LOT NUMBER FOR THE DISPOSABLE KIT THAT WAS IN USE. BAYER PRODUCT ANALYSIS TESTED A RETAINED SAMPLE FROM DISPOSABLE SJS-PT-J KIT, LOT NUMBER 8621072. FUNCTIONAL TESTING CONCLUDED THAT THE RETAINED DISPOSABLES PERFORMED TO SPECIFICATION WITH NO PROBLEMS OBSERVED. AN OFFER FOR ADDITIONAL APPLICATIONS TRAINING HAS BEEN MADE AND DECLINED BY THE SITE. THIS INFORMATION DOES NOT CONSTITUTE AN ADMISSION THAT THE DEVICE, THE COMPANY, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO A REPORTABLE EVENT.

Additional Manufacturer Narrative · 0

UDI RELATED DATA QUALITY UPDATES ONLY. CORRECTION: FOR UDI DATA DISCREPANCY/MISMATCH ON PREVIOUSLY SUBMITTED MDR AND TO ENSURE THE DEVICE IDENTIFICATION DATA IS IN ALIGNMENT WITH GUDID.

Description of Event or Problem · 0

BAYER MEDICAL CARE WAS NOTIFIED OF AN EXTRAVASATION THAT OCCURRED DURING AN ENHANCED CT SCAN WHILE A PATIENT WAS CONNECTED TO A MEDRAD® STELLANT CT INJECTION SYSTEM (SN (B)(6)). AN ESTIMATED 56 ML OF CONTRAST EXTRAVASATED AT THE INJECTION SITE. A COLD COMPRESS WAS APPLIED TO THE AFFECTED EXTREMITY AND AN ULTRASOUND EXAMINATION WAS PERFORMED THE FOLLOWING DAY WHICH CONFIRMED THAT THE ISSUE HAD RESOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1856770 STELLANT DUAL CT INJECTOR CT INJECTION SYSTEM DXT BAYER MEDICAL CARE INC. 85937286 00616258020860

Patients

Seq Age Sex Outcome Treatment
1 70 YR Female Other SJS-PT-J - LOT 8621072