ALINITY HQ PROCESSING MODULE

MDR report key: 17985443 · Received October 23, 2023

Report

Report Number
2919069-2023-00043
Event Type
Malfunction
Date Received
October 23, 2023
Date of Event
October 13, 2023
Report Date
June 26, 2024
Manufacturer
ABBOTT LABORATORIES
Product Code
GKZ
UDI-DI
00380740138851
PMA / PMN Number
K220031
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FIELD SERVICE REPRESENTATIVE (FSR) CHECKED OPTICS FOR LEAKAGE. S1 AND S2 TUBINGS FROM NOZZLE TO INJECTION MANIFOLD WERE REPLACED, AS WELL AS TUBING FROM THE INJECTION MANIFOLD TO THE WBC AND RBC STAGING PUMPS. CHECK VALVES FOR BUBBLE MIXING IN BOTH WBC CUPS AND TUBING IN DRAIN PINCH VALVES WERE REPLACED. THE CHECK VALVE AFTER V09 AND AFTER THE WASTE ASP/DISP CHAMBER WERE PROACTIVELY REPLACED. REPLACEMENT OF VALVES V55 AND V57 IN THE SAMPLING SWITCHING MANIFOLD IS WHAT RESOLVED THE ISSUE. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. A REVIEW OF CUSTOMER DATA ALIGNED WITH THE CUSTOMER¿S ISSUE AND NO ADDITIONAL ISSUES WERE IDENTIFIED. AN INSTRUMENT SERVICE HISTORY REVIEW REVEALED NO ADDITIONAL ERRATIC OR DISCREPANT PATIENT RESULTS REPORTED FOR (B)(4).THERE WERE NO SERVICE OR COMPLAINT ISSUES ON OR AROUND THE DATE THIS COMPLAINT WAS INITIATED THAT MAY HAVE CONTRIBUTED TO THIS ISSUE. A SEARCH FOR SIMILAR COMPLAINTS DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY. A REVIEW OF TRACKING AND TRENDING FOR THE ASSY, VALVE, LVM09, 2/2 N.C., 24VDC DID NOT IDENTIFY ANY TRENDS. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. BASED ON THE AVAILABLE INFORMATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY HQ PROCESSING MODULE FOR SERIAL (B)(6) OR THE ASSY, VALVE, LVM09, 2/2 N.C., 24VDC WAS IDENTIFIED.

Additional Manufacturer Narrative · 0

CORRECTION WAS MADE TO FIELD D2B PRO CODE. THE INCORRECT CODE OF GRZ WAS CHANGED TO GKZ. THIS WAS A TYPOGRAPHICAL ERROR CORRECTION AND THERE IS NO FURTHER IMPACT TO THE SUBMISSION. ADDITIONAL INFORMATION PROVIDED IN D4 PRIMARY UDI NUMBER:(B)(4).(B)(6).

Additional Manufacturer Narrative · 0

COMPLETED INFORMATION FOR SECTION A1 PATIENT IDENTIFIER: (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY DEPRESSED WBC RESULTS GENERATED ON THE ALINITY HQ PROCESSING MODULE ((B)(6)) FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6): (B)(6) RESULT = 8.96 K/UL, (B)(6) RESULT = 51.9 K/UL ADDITIONAL LABORATORY DATA WAS PROVIDED: (B)(6) GENERATED DATA INVALIDATING FLAGS FOR NEUTROPHILS, LYMPHOCYTES, EOSINOPHILS, BASOPHILS, AND IG DIFFERENTIAL RESULTS. THE DISCREPANT RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY DEPRESSED WBC RESULTS GENERATED ON THE ALINITY HQ PROCESSING MODULE (B)(6) FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) : (B)(6) RESULT = 8.96 K/UL, (B)(6) RESULT = 51.9 K/UL ADDITIONAL LABORATORY DATA WAS PROVIDED: (B)(6) GENERATED DATA INVALIDATING FLAGS FOR NEUTROPHILS, LYMPHOCYTES, EOSINOPHILS, BASOPHILS, AND IG DIFFERENTIAL RESULTS. THE DISCREPANT RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY DEPRESSED WBC RESULTS GENERATED ON THE ALINITY HQ PROCESSING MODULE (HQ00939) FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: SID 022823101330: HQ00939 RESULT = 8.96 K/UL, HQ00926 RESULT = 51.9 K/UL ADDITIONAL LABORATORY DATA WAS PROVIDED: HQ00939 GENERATED DATA INVALIDATING FLAGS FOR NEUTROPHILS, LYMPHOCYTES, EOSINOPHILS, BASOPHILS, AND IG DIFFERENTIAL RESULTS. THE DISCREPANT RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.