FDA Adverse Event Injury Summary report: N

CLICKFINE 32GX4MM DONGBAO 7CT

MDR report key: 17980445 · Received October 20, 2023

Report

Report Number
3002806818-2023-00019
Event Type
Injury
Date Received
October 20, 2023
Date of Event
August 24, 2023
Report Date
October 20, 2023
Manufacturer
YPSOMED AG
Product Code
FMI
PMA / PMN Number
K171984
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE NURSE'S FINGER BLED WHEN IT WAS PUNCTURED WHILE TAKING OFF THE OUTER PROTECTIVE CAP OF THE CLICKFINE DURING AN INSULIN INJECTION WITH A NOVOPEN. THE CANNULA WAS FOUND BENT AND STRETCHED OUT OF THE NEEDLE HUB. REVIEW OF MANUFACTURING DOCUMENTATION SHOWED NO ABNORMALITIES OR DEVIATIONS FROM THE VALIDATED MANUFACTURING PROCESS FOR THE MAIN BATCH 221515. ASSEMBLY ROCESS REVIEW OF THE STRAIGHTNESS CHECK FOR THE CANNULA IS FOUND TO BE 100% CONTROLLED BEFORE THE INNER PROTECTIVE CAP IS FITTED. NON-CONFORMING PARTS WERE AUTOMATICALLY SORTED OUT AS EXPECTED.

Description of Event or Problem · 0

THE NURSE'S FINGER BLED WHEN IT WAS PUNCTURED WHILE TAKING OFF THE OUTER PROTECTIVE CAP OF THE CLICKFINE DURING AN INSULIN INJECTION WITH A NOVOPEN. THE CANNULA WAS FOUND BENT AND STRETCHED OUT OF THE NEEDLE HUB.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1120830 CLICKFINE 32GX4MM DONGBAO 7CT PEN NEEDLE FMI YPSOMED AG CLICKFINE 32GX4MM DONGBAO 7CT 221515-02

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other