CLICKFINE 32GX4MM DONGBAO 7CT
Report
- Report Number
- 3002806818-2023-00019
- Event Type
- Injury
- Date Received
- October 20, 2023
- Date of Event
- August 24, 2023
- Report Date
- October 20, 2023
- Manufacturer
- YPSOMED AG
- Product Code
- FMI
- PMA / PMN Number
- K171984
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
THE NURSE'S FINGER BLED WHEN IT WAS PUNCTURED WHILE TAKING OFF THE OUTER PROTECTIVE CAP OF THE CLICKFINE DURING AN INSULIN INJECTION WITH A NOVOPEN. THE CANNULA WAS FOUND BENT AND STRETCHED OUT OF THE NEEDLE HUB. REVIEW OF MANUFACTURING DOCUMENTATION SHOWED NO ABNORMALITIES OR DEVIATIONS FROM THE VALIDATED MANUFACTURING PROCESS FOR THE MAIN BATCH 221515. ASSEMBLY ROCESS REVIEW OF THE STRAIGHTNESS CHECK FOR THE CANNULA IS FOUND TO BE 100% CONTROLLED BEFORE THE INNER PROTECTIVE CAP IS FITTED. NON-CONFORMING PARTS WERE AUTOMATICALLY SORTED OUT AS EXPECTED.
THE NURSE'S FINGER BLED WHEN IT WAS PUNCTURED WHILE TAKING OFF THE OUTER PROTECTIVE CAP OF THE CLICKFINE DURING AN INSULIN INJECTION WITH A NOVOPEN. THE CANNULA WAS FOUND BENT AND STRETCHED OUT OF THE NEEDLE HUB.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1120830 | CLICKFINE 32GX4MM DONGBAO 7CT | PEN NEEDLE | FMI | YPSOMED AG | CLICKFINE 32GX4MM DONGBAO 7CT | 221515-02 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |