FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 17978943 · Received October 20, 2023

Report

Report Number
1000113657-2023-00524
Event Type
Malfunction
Date Received
October 20, 2023
Date of Event
September 25, 2023
Report Date
October 20, 2023
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIPS TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-062: USER HAD POOR TECHNIQUE NOTE: MANUFACTURER CONTACTED CUSTOMER IN SEVERAL FOLLOW-UP CALLS TO ENSURE THE INITIAL CONCERN IS RESOLVED - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 240, 290, 320, 266 AND 333 MG/DL. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT IS 120 MG/DL AND EXPECTED 2 HOUR POST MEAL BLOOD GLUCOSE TEST RESULT IS 160 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULT OF 225 MG/DL USING TRUE METRIX METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE LIVING ROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 02/25/2025 AND OPEN VIAL DATE IS A FEW DAYS PRIOR TO CALL. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 240 MG/DL DATE: (B)(6) 2023 TIME: AM FASTING. RESULT 2: 290 MG/DL DATE: (B)(6) 2023 TIME: PM 2 HRS. AFTER MEAL. RESULT 3: 320 MG/DL DATE: (B)(6) 2023 TIME: PM 2 HRS. AFTER MEAL RESULT 4: 266 MG/DL DATE: (B)(6) 2023 TIME: AM FASTING. RESULT 5: 333 MG/DL DATE: (B)(6) 2023 TIME: AM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1908096 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TRUE METRIX 50CTMG/DL ZB5344S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1 Unknown