3M¿ RANGER¿ BLOOD/FLUID WARMING HIGH FLOW
Report
- Report Number
- 2110898-2023-00094
- Event Type
- Malfunction
- Date Received
- October 20, 2023
- Report Date
- June 6, 2024
- Manufacturer
- 3M COMPANY
- Product Code
- BSB
- UDI-DI
- 10707387792685
- PMA / PMN Number
- K973741
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO CORRECT SECTION D3 MANUFACTURER NAME, CITY AND STATE, FROM 3M HEALTH CARE TO 3M COMPANY.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO CORRECT SECTION D4 MODEL # FROM BLANK TO 24355, AND SECTION D4 UNIQUE IDENTIFIER (UDI) # FROM (B)(4) TO (B)(4).
H10: THE DEVICE HAS NOT BEEN RETURNED TO 3M FOR ANALYSIS. 3M IS UNABLE TO VERIFY THE LEAK, IDENTIFY EXACT LOCATION AND ROOT CAUSE WITHOUT THE SAMPLE AND LEAK LOCATION. POSSIBLE CAUSES OF LEAKS INCLUDE OVER-PRESSURIZATION ABOVE 300MMHG, INSERTION OR REMOVAL OF PRESSURIZED HEAT EXCHANGER FROM HEATING UNIT, AND HEAT EXCHANGER NOT FULLY INSERTED INTO THE HEATING UNIT. 3M WILL CONTINUE TO MONITOR.
A HOSPITAL ALLEGED WHILE USING 3M¿ RANGER¿ BLOOD/FLUID WARMING HIGH FLOW SET, 24355 THE HEAT EXCHANGER TORE INSIDE THE RANGER DURING AN EMERGENCY. NO PATIENT OR STAFF INJURY WAS ALLEGED, AND NO MEDICAL INTERVENTIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 933350 | 3M¿ RANGER¿ BLOOD/FLUID WARMING HIGH FLOW | BAIR HUGGER BLOOD/FLUID WARMER | BSB | 3M COMPANY | 24355 | HX9383 | 10707387792685 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |