PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 17976392 · Received October 20, 2023

Report

Report Number
2024168-2023-11615
Event Type
Injury
Date Received
October 20, 2023
Date of Event
September 12, 2023
Report Date
January 2, 2024
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED SUTURE SEPARATION WAS CONFIRMED AS A LINK SEPARATION. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO MANUFACTURING NONCONFORMITIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY, SUTURE/LINK PULLED UNTIL IT BREAKS DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO DESIGN, MANUFACTURE OR LABELING OF THE DEVICE. D4: LOT# WAS UPDATED FROM 2092141 TO 2101841. D4: EXPIRATION DATE UPDATED. H4: DEVICE MFG DATE UPDATED.

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.

Description of Event or Problem · 0

THIS WAS REPORTED AS AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY USING A PROGLIDE DEVICE AFTER A DIAGNOSTIC PROCEDURE WITH A 6F SHEATH. REPORTEDLY, THE SUTURE BROKE WHILE PULLING THE PLUNGER IN STEP 3. A NEW PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2038541 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 2101841

Patients

Seq Age Sex Outcome Treatment
1 80 YR Male Required Intervention