FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 2.0 TEST KIT 24T JAPAN

MDR report key: 17964664 · Received October 19, 2023

Report

Report Number
1221359-2023-01558
Event Type
Malfunction
Date Received
October 19, 2023
Date of Event
September 1, 2023
Report Date
October 19, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JP
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

B3: THE DATE PROVIDED IS AN APPROXIMATION AS THE EXACT EVENT DATE WAS NOT PROVIDED. (B)(4) THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 192-000J THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 192-000. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M232343 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 192-000J/ LOT: M232343, TEST BASE PART NUMBER 192-430/ LOT: M232343. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M232343 SHOWED THAT THE COMPLAINT RATE IS (B)(4) . ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS CROSS-CONTAMINATION. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON AN UNKNOWN DATE ON A NASAL SAMPLE. REPEAT TESTING WAS NOT PERFORMED. CONFIRMATION TESTING WAS PERFORMED VIA PCR (PLATFORM: UNKNOWN) ON A NASAL SAMPLE AND GENERATED A NEGATIVE RESULT. THE PATIENT WAS ASYMPTOMATIC. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1631612 ID NOW COVID-19 2.0 TEST KIT 24T JAPAN REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M232343

Patients

Seq Age Sex Outcome Treatment
1 Unknown