INPEN MMT-105NNBLNA NOVO NORDISK BLUE
Report
- Report Number
- 3012822846-2023-01720
- Event Type
- Malfunction
- Date Received
- October 19, 2023
- Date of Event
- August 2, 2023
- Report Date
- October 19, 2023
- Manufacturer
- COMPANION MEDICAL INC
- Product Code
- FMF
- UDI-DI
- 000010862088000344
- PMA / PMN Number
- K201337
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- 003
Narratives
CUSTOMER REPORTS: SCREW MOVEMENT ISSUE. PER VISUAL INSPECTION: NO PHYSICAL DAMAGE TO CARTRIDGE HOLDER OR INPEN FRONT AND BACK SHELL WAS NOTED. CARTRIDGE HOLDER LOCKS PROPERLY IN PLACE. INPEN PAIRED TO THE COMMERCIAL APP. INPEN RECEIVED WITH LEADSCREW 3/4 OF TRAVEL. PERFORMED DUST/DEBRIS INVESTIGATION AND FOUND VISIBLE HORIZONTAL ABRASIONS ON THE OUTSIDE OF ELECTRONICS HOUSING. THIS INDICATED DEBRIS WAS LODGED BETWEEN THE DOSE KNOB AND ELECTRONICS HOUSING CAUSING THE LEADSCREW ANOMALY. UNABLE TO PERFORM BASELINE/WIRELESS FUNCTIONALITY AND DISPLACEMENT DOSE ACCURACY TEST. PERFORMED LEADSCREW RESET TORQUE. INPEN WITHIN SPECIFICATION (CW: 0.40 OZF-IN AND CCW: 4.90 OZF-IN). INPEN PASSED FRONT CAP INVESTIGATION. IN CONCLUSION: DECONSTRUCTIVE ANALYSIS DEMONSTRATED VISIBLE HORIZONTAL ABRASIONS ON THE ELECTRONICS HOUSING WERE NOTED. THIS INDICATED DEBRIS WAS LODGED BETWEEN THE KNOB AND ELECTRONICS HOUSING CAUSING THE LEADSCREW TO RETRACT. THIS CAN AFFECT INSULIN DELIVERY. THEREFORE, LEADSCREW ANOMALY WAS CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED THAT THE SCREW WAS NOT MOVING AS INTENDED. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE THE USE OF THE INPEN AND IT WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 96924 | INPEN MMT-105NNBLNA NOVO NORDISK BLUE | SYRINGE, PISTON | FMF | COMPANION MEDICAL INC | MMT-105NNBLNA | B0768 | 000010862088000344 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Male |