FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® MULTIPLE SAMPLE LUER ADAPTER

MDR report key: 17959816 · Received October 18, 2023

Report

Report Number
1024879-2023-00727
Event Type
Malfunction
Date Received
October 18, 2023
Date of Event
September 22, 2023
Report Date
November 27, 2023
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
00382903673001
PMA / PMN Number
K991088
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: MATERIAL #: 367300. LOT/BATCH #: 3090608. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION, UTILIZING A BORESCOPE TO SHINE LIGHT THROUGH THE CANNULA, AND NO ISSUES WERE OBSERVED RELATING TO INSUFFICIENT FLOW AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE INSUFFICIENT BLOOD FLOW. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Additional Manufacturer Narrative · 0

H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH THE BD VACUTAINER® MULTIPLE SAMPLE LUER ADAPTER, THERE WAS INSUFFICIENT BLOOD FLOW. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH: AT THE TIME OF THE BLOOD TEST THE BLOOD DOES NOT COME, TEST IMPOSSIBLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH THE BD VACUTAINER® MULTIPLE SAMPLE LUER ADAPTER, THERE WAS INSUFFICIENT BLOOD FLOW. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH: AT THE TIME OF THE BLOOD TEST THE BLOOD DOES NOT COME, TEST IMPOSSIBLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
175813 BD VACUTAINER® MULTIPLE SAMPLE LUER ADAPTER BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) 3090608 00382903673001

Patients

Seq Age Sex Outcome Treatment
1 Unknown