FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 17958730 · Received October 18, 2023

Report

Report Number
1221359-2023-01556
Event Type
Malfunction
Date Received
October 18, 2023
Date of Event
September 26, 2023
Report Date
October 18, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

FDA UDI (B)(4). TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 211557 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 211557 AND TEST BASE PART NUMBER 195-430H / LOT 206721. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 211557 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD HAVE POSSIBLY BEEN RELATED TO THE SPECIFIC PATIENT SAMPLE. H3 OTHER TEXT : SINGLE-USE: DEVICE DISCARDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2023 ON A NASAL SAMPLE. REPEAT TESTING WAS NOT PERFORMED. PCR CONFIRMATION TESTING (PLATFORM UNKNOWN) WAS PERFORMED ON THE SAME DAY AND GENERATED A NEGATIVE RESULT. CONSUMER HAD TAKEN A COVID-19 TEST (BRAND UNKNOWN) A FEW DAYS PRIOR AND GENERATED A NEGATIVE RESULT. THE CONSUMER WAS SYMPTOMATIC PRIOR TO TESTING. THE CONSUMER CONFIRMED THERE WAS NO HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CONSUMER CONFIRMED THERE WAS NO DELAY OR IMPACT TO THEIR TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
179460 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 211557 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 Female