FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 17957641 · Received October 18, 2023

Report

Report Number
1645337-2023-12233
Event Type
Injury
Date Received
October 18, 2023
Date of Event
January 18, 2023
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001270
PMA / PMN Number
P990075
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON OCTOBER 24, 2023, MENTOR BECAME AWARE THAT INCORRECT MANUFACTURING RECORD EVALUATION (MRE) STATEMENT WAS INADVERTENTLY SUBMITTED PREVIOUSLY SINCE THE LOT NUMBER IS UNKNOWN. THE CORRECT STATEMENT IS AS FOLLOWS: "SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION COULD BE PERFORMED." MANUFACTURER¿S REFERENCE NUMBER: (B)(4), INCORRECT MRE STATEMENT SUBMITTED.

Additional Manufacturer Narrative · 0

ON NOVEMBER 27, 2023, DEVICE EVALUATION WAS COMPLETED. DEVICE EVALUATION SUMMARY: MENTOR CONDUCTED A VISUAL INSPECTION, MICROSCOPIC EXAMINATION, AND THICKNESS MEASUREMENT OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE SAL SMOOTH RND DIAP 375CC BREAST IMPLANT WAS FOUND TO HAVE AN AREA OF MISSING MATERIAL MEASURING 7.5 X 3.0 CM ON THE POSTERIOR VIEW. MICROSCOPIC EXAMINATION WAS PERFORMED AND THE CAUSE OF THE DEFLATION COULD NOT BE IDENTIFIED. THEREFORE A THICKNESS MEASUREMENT WAS CONDUCTED ON THE SHELL AT THE TEAR, AND THE THICKNESS WAS WITHIN MANUFACTURING SPECIFICATIONS. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. ALTHOUGH NO CONCLUSION COULD BE REACH ON THE CAUSE OF THE REPORTED EVENT, THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING CAUTION: CAUSES OF DEFLATION OF SALINE IMPLANTS INCLUDE BUT ARE NOT LIMITED TO THE FOLLOWING EVENTS: CREASE FOLD FAILURE, INTRAOPERATIVE OR POSTOPERATIVE TRAUMA, EXCESSIVE STRESSES OR MANIPULATIONS AS MAY OCCUR DURING NORMAL, DAILY ROUTINES INCLUDING CUSTOMARY AND PURPOSEFUL TRAUMA SUCH AS THAT WHICH CAN OCCUR DURING VIGOROUS EXERCISE, ATHLETICS, ETC. A SECOND PRODUCT WAS RECEIVED LOT 5925738. NO ADVERSE EVENTS WERE REPORTED FOR THIS CONCOMITANT (CONTRALATERAL) DEVICE, THEREFORE NO FURTHER ANALYSIS IS REQUIRED. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RIGHT DEFLATION. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON NOVEMBER 7, 2023, MENTOR BECAME AWARE REGARDING THE DEVICE INFORMATION. HENCE, THE FOLLOWING FIELDS HAVE BEEN UPDATED ON THIS FORM: FIELD D1: FOR BRAND NAME HAS BEEN UPDATED TO "MENTOR SMOOTH ROUND MODERATE PROFILE." FIELD D4: FOR CATALOG NUMBER HAS BEEN UPDATED TO "3501660." FIELD D4: FOR LOT NUMBER HAS BEEN UPDATED TO "5925738." FIELD D4: FOR UNIQUE IDENTIFIER( UDI) HAS BEEN UPDATED TO "(B)(4)." A MANUFACTURING RECORD EVALUATION IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ON NOVEMBER 22, 2023, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH UNKNOWN SIZE UNKNOWN SALINE IMPLANTS AND EXPERIENCED BREAST IMPLANT DEFLATION ON HER RIGHT SIDE POSTOPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT SURGERY ON (B)(6) 2023.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
609580 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 5925738 00081317001270

Patients

Seq Age Sex Outcome Treatment
1 37 YR Female Required Intervention