FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL TIP 1ML

MDR report key: 17956855 · Received October 18, 2023

Report

Report Number
1911916-2023-00783
Event Type
Malfunction
Date Received
October 18, 2023
Date of Event
October 12, 2023
Report Date
October 26, 2023
Manufacturer
BECTON DICKINSON
Product Code
FMF
UDI-DI
30382903096535
PMA / PMN Number
K110771
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
PHARMACIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. DEVICE PROBLEM CODE: A020102 - NONSTANDARD DEVICE. PATIENT PROBLEM CODE: F27 ¿ NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative · 0

PR 9081353 FOLLOW UP REPORT IT WAS REPORTED BY THE CONSUMER THE SYRINGES ARE MISSING PRINTING LINES AND NUMBERS. AS A SAMPLE WAS NOT RETURNED, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309653, LOT 3208287. THERE WAS DOCUMENTATION OF THIS TYPE OF DEFECT DURING THE PRODUCTION RUN OF THIS LOT. A QUALITY NOTIFICATION WAS ISSUED FOR THE SCALE MARKING ISSUE. IT COULD BE POSSIBLE THE CUSTOMER RECEIVED ESCAPES FROM THIS INCIDENT EXPERIENCED DURING PRODUCTION. BASED ON THE INVESTIGATION AND WITH NO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED. H3 OTHER TEXT : SEE H10 NARRATIVE.

Description of Event or Problem · 0

MATERIAL#:309653. BATCH#:3208287. IT WAS REPORTED BY CUSTOMER THAT THEY HAVE NOTED THAT ON SOME OF THE BD 50ML SYRINGES NDC 08290-3096-80 THAT THE PRINTING OF THE LINES AND NUMBERS ARE MISSING IN SPOTS AND THEY HAVE HAD TO THROW THEM AWAY. TODAY THEY THREW OUT AT LEAST 30 AND THEY ALL HAD THE SAME LOT # WHICH IS 3208287 EXP DATE 6/30/2028. THEY HAVE NOT USED ANY OF THE SYRINGES THAT WERE NOTED TO HAVE MISSING LINES ORE NUMBERS, BUT WANTED TO REPORT THIS TO BD. VERBATIM: CUSTOMER STATED THAT WE HAVE NOTED THAT ON SOME OF THE BD 50ML SYRINGES NDC 08290-3096-80 THAT THE PRINTING OF THE LINES AND NUMBERS ARE MISSING IN SPOTS AND WE HAVE HAD TO THROW THEM AWAY. TODAY WE THREW OUT AT LEAST 30 AND THEY ALL HAD THE SAME LOT # WHICH IS 3208287 EXP DATE 6/30/2028. WE HAVE NOT USED ANY OF THE SYRINGES THAT WERE NOTED TO HAVE MISSING LINES ORE NUMBERS, BUT WANTED TO REPORT THIS TO YOUR COMPANY SO YOU CAN BE AWARE.

Description of Event or Problem · 0

PR 9081353 UPDATE - ADDITIONAL INFORMATION RECEIVED MATERIAL#:309653 BATCH#:3208287 IT WAS REPORTED BY CUSTOMER THAT THEY HAVE NOTED THAT ON SOME OF THE BD 50ML SYRINGES NDC 08290-3096-80 THAT THE PRINTING OF THE LINES AND NUMBERS ARE MISSING IN SPOTS AND THEY HAVE HAD TO THROW THEM AWAY. TODAY THEY THREW OUT AT LEAST 30 AND THEY ALL HAD THE SAME LOT # WHICH IS 3208287 EXP DATE 6/30/2028. THEY HAVE NOT USED ANY OF THE SYRINGES THAT WERE NOTED TO HAVE MISSING LINES ORE NUMBERS, BUT WANTED TO REPORT THIS TO BD. VERBATIM: CUSTOMER STATED THAT WE HAVE NOTED THAT ON SOME OF THE BD 50ML SYRINGES NDC 08290-3096-80 THAT THE PRINTING OF THE LINES AND NUMBERS ARE MISSING IN SPOTS AND WE HAVE HAD TO THROW THEM AWAY. TODAY WE THREW OUT AT LEAST 30 AND THEY ALL HAD THE SAME LOT # WHICH IS 3208287 EXP DATE 6/30/2028. WE HAVE NOT USED ANY OF THE SYRINGES THAT WERE NOTED TO HAVE MISSING LINES ORE NUMBERS, BUT WANTED TO REPORT THIS TO YOUR COMPANY SO YOU CAN BE AWARE. ADDITIONAL INFO FROM CUSTOMER: -WHAT IS THE TOTAL QUANTITY OF PRODUCTS FOUND DEFECTIVE? ~30 -IS THERE ANY ADVERSE EVENT OR SERIOUS INJURY OCCURRED? NO -WHAT IS THE EVENT DATE? 10/12/2023 -ARE YOU ABLE TO PROVIDE SAMPLE TO BD FOR INVESTIGATION? IF NO, CAN A PHOTO BE PROVIDED? IF YES, PLEASE PROVIDE MAILING ADDRESS. PRODUCTS WERE DISPOSED OF PRIOR TO STAFF TELLING ME. I CAN NOT PROVIDE A SAMPLE OR PHOTO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
539503 SYRINGE 50ML LL TIP 1ML SYRINGE, PISTON FMF BECTON DICKINSON 3208287 30382903096535

Patients

Seq Age Sex Outcome Treatment
1 Unknown