FDA Adverse Event
Malfunction
Summary report: N
FOCUS RTP SYSTEM
MDR report key: 179500
·
Received July 22, 1998
Report
- Report Number
- 1937649-1998-00005
- Event Type
- Malfunction
- Date Received
- July 22, 1998
- Date of Event
- July 10, 1998
- Report Date
- July 22, 1998
- Manufacturer
- COMPUTERIZED MEDICAL SYSTEMS
- Product Code
- MUJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIST
Narratives
Description of Event or Problem · 1
WHEN USING A MULTILEAF COLLIMATOR TO FORM A PORT, & CALCULATING DOSE USING THE "FFT" CONVOLUTION ALGORITHM, THE FOCUS RTP SYSTEM IS NOT LOOKING AT THE MULTILEAF COLLIMATOR LEAF POSITIONS TO DEFINE THE PORTAL OUTLINE USED IN DOSE CALCULATION. RATHER, FOCUS IS USING THE ANATOMICAL STRUCTURE OR CUSTOMIZED PORT OUTLINE. THE RESULT COULD BE AN INNOCENT ISODOSE DISTRIBUTION FOR THE CONDITION, WHERE THE MULTILEAF COLLIMATOR PORTAL OUTLINE DOES NOT COINCIDE WITH THE ANATOMICAL STRUCTURE OR CUSTOMIZED PORT OUTLINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FOCUS RTP SYSTEM | RADIATION TREATMENT PLANNING SYSTEM | MUJ | COMPUTERIZED MEDICAL SYSTEMS | 9200 | RELEASE 1.5.0 & HIGHER |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |