FDA Adverse Event Malfunction Summary report: N

FOCUS RTP SYSTEM

MDR report key: 179500 · Received July 22, 1998

Report

Report Number
1937649-1998-00005
Event Type
Malfunction
Date Received
July 22, 1998
Date of Event
July 10, 1998
Report Date
July 22, 1998
Manufacturer
COMPUTERIZED MEDICAL SYSTEMS
Product Code
MUJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIST

Narratives

Description of Event or Problem · 1

WHEN USING A MULTILEAF COLLIMATOR TO FORM A PORT, & CALCULATING DOSE USING THE "FFT" CONVOLUTION ALGORITHM, THE FOCUS RTP SYSTEM IS NOT LOOKING AT THE MULTILEAF COLLIMATOR LEAF POSITIONS TO DEFINE THE PORTAL OUTLINE USED IN DOSE CALCULATION. RATHER, FOCUS IS USING THE ANATOMICAL STRUCTURE OR CUSTOMIZED PORT OUTLINE. THE RESULT COULD BE AN INNOCENT ISODOSE DISTRIBUTION FOR THE CONDITION, WHERE THE MULTILEAF COLLIMATOR PORTAL OUTLINE DOES NOT COINCIDE WITH THE ANATOMICAL STRUCTURE OR CUSTOMIZED PORT OUTLINE.