FDA Adverse Event Malfunction Summary report: N

INPEN MMT-105ELBLNA ELI LILY BLUE

MDR report key: 17947895 · Received October 17, 2023

Report

Report Number
3012822846-2023-01699
Event Type
Malfunction
Date Received
October 17, 2023
Date of Event
September 20, 2023
Report Date
October 17, 2023
Manufacturer
COMPANION MEDICAL INC
Product Code
FMF
UDI-DI
000010862088000313
PMA / PMN Number
K201337
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INPEN PAIRED WITH COMMERCIAL MOBILE APP WITH SMALL DOSE. PERFORMED ENCODER BOND INVESTIGATION AND FOUND THAT THE ENCODER PATTERN WHEEL TABS ROTATING AND TRAVELING OFF THE KEYED SLOTS OF DOSE NUT GUIDES. ENCODER PATTERN WHEEL SHOULD NEVER ROTATE. THIS CAUSES AN UNEXPECTED TRAVEL OF THE ENCODER PATTERN WHEEL CREATING RESISTANCE TO DIAL AND DISPENSING. ABRASIONS DOWN THE LENGTH OF THE DOSE NUT. UNABLE TO PERFORM BASELINE/WIRELESS FUNCTIONALITY AND DISPLACEMENT DOSE ACCURACY TEST. PERFORMED LEADSCREW RESET TORQUE TEST. INPEN WITHIN SPECIFICATION (CW 0.70 OZF-IN AND CCW: 4.35 OZF-IN). INPEN FAILED FRONT CAP INVESTIGATION. INPEN FRONT SHELL DOES NOT FIT SECURELY ONTO CARTRIDGE HOLDER DUE TO SNAP ARM BEING CRACKED / BROKEN. IN CONCLUSION: IT WAS DETERMINED FROM DESTRUCTIVE ANALYSIS THAT THE DIFFICULT TO DIAL/DOSE WAS CAUSED BY A PATTERN WHEEL MISALIGNMENT DUE TO AN ENCODER BASE BOND FAILURE. THIS CAN AFFECT INSULIN DELIVERY. THEREFORE, THE CUSTOMER CONCERN OF DOSE DIAL BEING DIFFICULT TO DIAL/DOSE WAS CONFIRMED. CAP ANOMALY WAS CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED RETRACTING THE INPEN SCREW WAS DIFFICULT AND DOSE KNOB/DIAL DIFFICULT TO TURN. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INPEN WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
942049 INPEN MMT-105ELBLNA ELI LILY BLUE SYRINGE, PISTON FMF COMPANION MEDICAL INC MMT-105ELBLNA B0802 000010862088000313

Patients

Seq Age Sex Outcome Treatment
1 53 YR Male