FDA Adverse Event Malfunction Summary report: N

NONE

MDR report key: 17946667 · Received October 16, 2023

Report

Report Number
2955842-2023-19259
Event Type
Malfunction
Date Received
October 16, 2023
Date of Event
September 19, 2023
Report Date
September 19, 2023
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874112205
PMA / PMN Number
K171426
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. THE REPORTED EVENT WAS ADDRESSED WITH PHONE SUPPORT. THE CALLER REMOVED THE ENDOSCOPE AND ERASED THE GUIDED TOOL CHANGE FEATURE. THE SCOPE WAS REINSTALLED, AND THE IMAGE WAS NOW APPEARING NORMALLY. NO SITE VISIT WAS CONDUCTED. THE SYSTEM WAS WORKING PROPERLY, AND NO ADDITIONAL ACTION WAS REQUIRED AS THE ISSUE WAS RESOLVED. SIMILAR ISSUE WAS NOTED UNDER PATIENT IDENTIFIER# (B)(6) FOR THE 30-DEGREE ENDOSCOPE. A RETURN MATERIAL AUTHORIZATION (RMA) WAS ISSUED TO THE CUSTOMER REQUESTING TO RETURN THE AFFECTED ENDOSCOPE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PROSTATECTOMY SURGICAL PROCEDURE, THE IMAGE WAS UPSIDE DOWN. THE SITE REMOVED THE ENDOSCOPE AND ERASED THE GUIDED TOOL CHANGE FEATURE. THE ENDOSCOPE WAS REINSTALLED, AND THE IMAGE WAS NOW APPEARING NORMALLY. THE SURGEON IS PROCEEDING WITH THE CASE. THE SYSTEM LOGS WERE REVIEWED WITH NO ISSUES FOUND. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER ON AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: SURGEON CONFIRMED THIS HAPPENED IN THE MIDDLE OF THE CASE. SURGEON DOESN'T RECALL THE EXACT PORTION BUT DOES RECALL THAT BOTH THE 30-DEGREE SCOPE AND THE 0-DEGREE HAD THIS ISSUE. THE ENDOSCOPES WERE INSERTED PROPERLY WITH PROPER ORIENTATION. THE IMAGE ORIENTATION APPEARED CORRECT IN THE CONSOLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
283504 NONE ENDOSCOPE NAY INTUITIVE SURGICAL, INC 470026-62 N/A 00886874112205

Patients

Seq Age Sex Outcome Treatment
1 Male DA VINCI INSTRUMENTS AND ACCESSORIES