FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1793983 · Received August 11, 2010

Report

Report Number
2649622-2010-07244
Event Type
Malfunction
Date Received
August 11, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED, AND VARYING RESISTANCE/IMPEDANCE WAS OBSERVED. THE MAX RV IMPEDANCE MEASUREMENT VARIED OVER THE DURATION OF THE RECORD WITH A RANGE FROM 696 - 1392 OHMS.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED, AND VARYING RESISTANCE/IMPEDANCE WAS OBSERVED. THE MAX RV IMPEDANCE MEASUREMENT VARIED OVER THE DURATION OF THE RECORD WITH A RANGE FROM 696 - 1392 OHMS. (B)(4) THE DEVICE WAS RETURNED, ANALYZED, AND ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. DEVICE MEETS EXPECTED LONGEVITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RV PACING IMPEDANCE WAS VARYING FROM 600 OHMS UP TO 1300 OHMS ("SPIKING") THROUGHOUT THE LONG TERM TREND. THE COMPANY TECHNICAL REPRESENTATIVE SUGGESTED THE DATA ARE CONSISTENT WITH EITHER LEAD FRACTURE OR A CONNECTOR ISSUE. THE LEAD AND DEVICE REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 86 YR Other 5076 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD