FDA Adverse Event Malfunction Summary report: N

BD SURGIPHOR¿ STERILE WOUND IRRIGATION SYSTEM

MDR report key: 17939661 · Received October 16, 2023

Report

Report Number
1423507-2023-00102
Event Type
Malfunction
Date Received
October 16, 2023
Date of Event
August 3, 2023
Report Date
March 11, 2024
Manufacturer
CAREFUSION, INC
Product Code
FQH
UDI-DI
00382909101102
PMA / PMN Number
K221504
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) INITIAL MDR. A FOLLOW UP WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. --- SECTION D. 2B PROCODE: FQH (LAVAGE, JET) AND FRO (DRESSING, WOUND, DRUG).

Additional Manufacturer Narrative · 0

(B)(4)SUPPLEMENTAL MDR. ONE PHOTO WAS RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. FROM THE PHOTO, THE BOTTLE WAS OBSERVED TO BE CRACKED AROUND THE TOP CAP AREA CONFIRMING THE CUSTOMER¿S EXPERIENCE.A DEVICE HISTORY RECORD COULD NOT BE EVALUATED AS THE LOT NUMBER IS UNKNOWN. THE MOST PROBABLE ROOT CAUSE CAN BE ATTRIBUTED TO POST MANUFACTURING HANDLING. DURING SHIPMENT, BOXES MAY BE DROPPED OR IMPACTED WHICH CAN BE EXCESSIVE HANDLING CAUSING THE LEAKAGE. THIS INCIDENT HAS BEEN ADDED TO OUR DATABASE OF REPORTED INCIDENTS. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. ---- SECTION D. 2B PROCODE: FQH (LAVAGE, JET) AND FRO (DRESSING, WOUND, DRUG) H3 OTHER TEXT : SEE NARRATIVE BELOW

Description of Event or Problem · 0

THE SURGIPHOR BOTTLE EXPLODED WHILE THE CAP WAS BEING SCREWED ON.

Description of Event or Problem · 0

THE SURGIPHOR BOTTLE EXPLODED WHILE THE CAP WAS BEING SCREWED ON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
704200 BD SURGIPHOR¿ STERILE WOUND IRRIGATION SYSTEM SURGIPHOR FQH CAREFUSION, INC UNKNOWN 00382909101102

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown