FDA Adverse Event Malfunction Summary report: N

DA VINCI

MDR report key: 17936726 · Received October 14, 2023

Report

Report Number
2955842-2023-19219
Event Type
Malfunction
Date Received
October 14, 2023
Date of Event
September 19, 2023
Report Date
September 19, 2023
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874112212
PMA / PMN Number
K171426
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. THE REPORTED EVENT WAS ADDRESSED WITH PHONE SUPPORT. THE CALLER REMOVED THE ENDOSCOPE AND ERASED THE GUIDED TOOL CHANGE FEATURE. THE ENDOSCOPE WAS REINSTALLED, AND THE IMAGE WAS NOW APPEARING NORMALLY. NO SITE VISIT WAS CONDUCTED. THE SYSTEM WAS WORKING PROPERLY, AND NO ADDITIONAL ACTION WAS REQUIRED AS THE ISSUE WAS RESOLVED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PROSTATECTOMY SURGICAL PROCEDURE, THE IMAGE WAS UPSIDE DOWN. THE CALLER REMOVED THE ENDOSCOPE AND ERASED THE GUIDED TOOL CHANGE FEATURE. THE ENDOSCOPE WAS REINSTALLED, AND THE IMAGE WAS NOW APPEARING NORMALLY. THE SURGEON WAS PROCEEDING WITH THE CASE. THE SYSTEM LOGS WERE REVIEWED WITH NO ISSUES FOUND. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE SURGEON CONFIRMED THIS HAPPENED IN THE MIDDLE OF THE CASE. THE SURGEON DID NOT RECALL THE EXACT PORTION OF THE PROCEDURE BUT DID RECALL THAT BOTH THE 30-DEGREE AND THE 0-DEGREE HAD THIS ISSUE. THE ENDOSCOPES WERE INSERTED PROPERLY WITH PROPER ORIENTATION. THE IMAGE ORIENTATION APPEARED CORRECT IN THE CONSOLE. THE SURGEON WAS SURE THE ENDOSCOPES WERE AVAILABLE FOR RETURN. THE ISSUE WAS QUICKLY RESOLVED DURING THE CASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1716331 DA VINCI ENDOSCOPE NAY INTUITIVE SURGICAL, INC 470027-64 N/A 00886874112212

Patients

Seq Age Sex Outcome Treatment
1 Male DA VINCI INSTRUMENTS AND ACCESSORIES.