FDA Adverse Event Injury Summary report: N

IMP,TSV,4.7,11.5,MTX,MG

MDR report key: 17930488 · Received October 13, 2023

Report

Report Number
0002023141-2023-02893
Event Type
Injury
Date Received
October 13, 2023
Date of Event
July 11, 2023
Report Date
March 24, 2024
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024020009
PMA / PMN Number
K101977
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: EVALUATION METHOD CODES WERE ADDED: 3331, 4110. H10: NARRATIVE/DATA WAS UPDATED. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1250932. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. STERILIZATION RECORD OP# 150 ¿ STR2 WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1250932 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMVIE CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION. MONTHLY TRENDING TO DATE HAS NOT IDENTIFIED ANY STATISTICAL TRIGGERS SUGGESTING POTENTIAL DESIGN OR MANUFACTURING RELATED ISSUES. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THE INFECTION OF AN IMPLANT HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES IMPACTING THE MANUFACTURING AND STERILIZATION PROCESSES. THEREFORE, ESCALATION TO CAPA IS NOT REQUIRED.

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). G4: ADDITIONAL PMA/510(K) NUMBER ¿ K101880. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR INFECTION EVENTS IDENTIFYING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO THE WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED) FACTORS POTENTIALLY IMPACTING THE STERILITY OF THE IMPLANT AND ITS ENVIRONMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE INFECTION EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED. H3 OTHER TEXT : SUMMARY INVESTIGATION.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS AN INFECTION AROUND THE IMPLANT AT TOOTH LOCATION #11; SUSPECTED TO BE DUE TO THE FRACTURED PART THAT BECAME A FOREIGN BODY REACTION IN THE GINGIVA. SYMPTOMS AS A RESULT OF THE EVENT: PAIN, INFLAMMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1551721 IMP,TSV,4.7,11.5,MTX,MG DENTAL IMPLANT DZE ZIMMER DENTAL 1250932 00889024020009

Patients

Seq Age Sex Outcome Treatment
1 65 YR Female Required Intervention