FDA Adverse Event Injury Summary report: N

SELECT SECURE

MDR report key: 1792580 · Received August 11, 2010

Report

Report Number
2649622-2010-07042
Event Type
Injury
Date Received
August 11, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P030036
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. (B)(4) ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THE LEAD AND DEVICE WERE REMOVED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE DEVICE WAS RETURNED, ANALYZED AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. ADDITIONAL INFORMATION RECEIVED DURING FOLLOW-UP INDICATED THAT NO INFECTION WAS FOUND. THE DEVICE HAD REACHED ERI (ELECTIVE REPLACEMENT), PATIENT'S EF HAD IMPROVED, AND THE PATIENT NO LONGER WANTED DEVICE BECAUSE HE HAD RECEIVED TWO SHOCKS FROM AF WITH RVR. NO PLANS TO REIMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THE LEAD AND DEVICE WERE REMOVED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE DEVICE WAS RETURNED, ANALYZED AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SELECT SECURE IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 3830 ASKU

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention 6937A IMPLANTABLE TACHY LEAD| 6949 IMPLANTABLE TACHY LEAD