FDA Adverse Event
Injury
Summary report: N
SIGMA 100 VVI
MDR report key: 1792547
·
Received August 11, 2010
Report
- Report Number
- 6000024-2010-00045
- Event Type
- Injury
- Date Received
- August 11, 2010
- Date of Event
- March 9, 2010
- Manufacturer
- MEDTRONIC OF CANADA LTD.
- Product Code
- DXY
- PMA / PMN Number
- P980035/S2
- Removal / Correction Number
- Z-1510/1519-2009
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. THE EVENT OCCURRED OUTSIDE THE US. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) PRELIMINARY TESTING REVEALED NO OUTPUT AND NO TELEMETRY. THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.
Description of Event or Problem · 1
THE DEVICE WAS REPLACED PREVENTIVELY DUE TO THE FIELD ADVISORY. THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SIGMA 100 VVI | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC OF CANADA LTD. | SVVI103 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Other |