FDA Adverse Event Malfunction Summary report: N

BRILLIANT S+ VDD

MDR report key: 1792510 · Received August 11, 2010

Report

Report Number
2649622-2010-07132
Event Type
Malfunction
Date Received
August 11, 2010
Date of Event
April 4, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P890003/S50
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND. FULL LEAD RETURNED AND ANALYZED. THE LEAD WAS RECEIVED WITH THE STYLET STILL IN THE LEAD. AFTER REMOVING THE STYLET A KINK WAS OBSERVED ON THE STYLET. IT CAN'T BE DETERMINED WHEN THIS OCCURRED OR IF THAT IS WHY THE LEAD GOT STUCK IN THE 9FR INTRODUCER SHEATH.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS AN ATTEMPT TO IMPLANT THE VDD PACING THROUGH A 9FR INTRODUCER, AS LABELED. THE LEAD GOT STUCK IN THE INTRODUCER. WHEN IT WAS PULLED BACK OUT OF THE INTRODUCER, THE PHYSICIAN SAW A KINK IN THE DISTAL TIP. THE LEAD WAS RETURNED. ADDITIONAL COMMUNICATION FROM THE CUSTOMER CLARIFIED THAT THE ISSUE WAS NOT WITH THE VDD PACING LEAD, BUT RATHER THE 9 FR INTRODUCER WHICH REPORTEDLY "KINKED" INSIDE THE VEIN AND WOULD NOT ALLOW THE VDD LEAD TO PASS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.;

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS AN ATTEMPT TO IMPLANT THE VDD PACING THROUGH A 9FR INTRODUCER, AS LABELED. THE LEAD GOT STUCK IN THE INTRODUCER. WHEN IT WAS PULLED BACK OUT OF THE INTRODUCER, THE PHYSICIAN SAW A KINK IN THE DISTAL TIP. THE LEAD WAS NOT USED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BRILLIANT S+ VDD IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. IMW15Q ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other| R