FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1792507 · Received August 11, 2010

Report

Report Number
2649622-2010-07127
Event Type
Malfunction
Date Received
August 11, 2010
Date of Event
April 30, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) DISTAL CONDUCTOR DISTORTED; FULL LEAD WAS RETURNED FOR ANALYSIS. THE LEAD WAS RETURNED WITH THE HELIX EXTENDED. THE HELIX WILL NOT EXTEND DUE TO THE DISTORTED DISTAL COIL IN THE CONNECTOR. THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR AND BLOOD IN/ON HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE, THE LEAD WAS UNABLE TO FIX SINCE THE HELIX WOULD NOT EXTEND. THE LEAD WAS NOT IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention