FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1792448 · Received August 11, 2010

Report

Report Number
2649622-2010-07118
Event Type
Malfunction
Date Received
August 11, 2010
Date of Event
May 6, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MODE FOR THE LAST MONTH HAS NOT USED ATRIAL LEAD. PRIOR TO CHANGING THE MODE, THE ATRIAL LEAD IMPEDANCES WAS 3000 - 4000 FREQUENTLY. IMPEDANCE IS CURRENTLY AT 500 - 600 OHMS. OVERSENSING WAS OBSERVED. THE LEAD IS STILL IN USE AND THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 68 YR Other 5076 IMPLANTABLE PACING LEAD| ADDR01 IMPLANTABLE PULSE GENERATOR