FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1792359 · Received August 11, 2010

Report

Report Number
2649622-2010-07090
Event Type
Malfunction
Date Received
August 11, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. PERFORMANCE DATA COLLECTED FROM THE DEVICE SHOWED IMPEDANCE TRENDS WERE STEADY. ALSO NOISE WAS OBSERVED. THE LIFETIME VENTRICULAR SENSING INTEGRITY COUNTER IS 614 AND THE RATE HAS BEEN STEADY OVER THE IMPLANT TIME. A HIGH VALUE OF THIS COUNT CAN INDICATE A POSSIBLE INTERMITTENCY IN THE SYSTEM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAS RECORDED SHORT INTERVAL COUNT (SIC) OF 489 SINCE THE LAST FOLLOW-UP. THE TOTAL SIC FOR THE DEVICE IS 614. THE LEAD REMAINS IMPLANTED AND ACTIVE. THERE WERE NO PATIENT COMPLAINTS OR COMPLICATIONS REPORTED DUE TO THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 84 YR Other 5076 IMPLANTABLE PACING LEAD| D284DRG IMPLANTABLE PACEMAKER/CARDIO/DEFIB