FDA Adverse Event
Malfunction
Summary report: N
ADAPTA DR
MDR report key: 1792319
·
Received August 11, 2010
Report
- Report Number
- 6000144-2010-03218
- Event Type
- Malfunction
- Date Received
- August 11, 2010
- Date of Event
- May 5, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NVZ
- PMA / PMN Number
- P980035/S043
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS INTERROGATED PRIOR TO IMPLANT AND RECEIVED MESSAGE OF 'REPLACE PACER' THAT COULD NOT BE RESET. THE BATTERY VOLTAGE WAS 2.77V AND 408 OHMS. THE DEVICE WAS NOT IMPLANTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THHIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADAPTA DR | IMPLANTABLE PULSE GENERATOR | NVZ | MEDTRONIC MED REL, INC. | ADDRS1 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |