FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1792285 · Received August 10, 2010

Report

Report Number
2649622-2010-07060
Event Type
Injury
Date Received
August 10, 2010
Date of Event
March 31, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A ROUTINE CHECK UP ONE WEEK POST IMPLANT THERE WAS INCREASED THRESHOLD AND DECREASED SENSING. THIS WAS DETERMINED TO BE MICRO DISLODGEMENT OF THE RIGHT VENTRICULAR LEAD. THE LEAD WAS REPOSITIONED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 59 YR Required Intervention (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4076 IMPLANTABLE PACING LEAD