FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL
MDR report key: 1792178
·
Received August 10, 2010
Report
- Report Number
- 6000001-2010-02465
- Event Type
- Malfunction
- Date Received
- August 10, 2010
- Date of Event
- July 20, 2010
- Report Date
- July 20, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Removal / Correction Number
- 6000001-3/15/05-007-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE ROOT CAUSE FOR THIS ISSUE IS CURRENTLY BEING INVESTIGATED THROUGH CAPA (CORRECTIVE AND PREVENTIVE ACTION) INVESTIGATION MDQ-CAPA-(B)(4).
Additional Manufacturer Narrative · 1
(B)(4). THE ASSIGNABLE CAUSE WAS THE AIL PCB WAS OUT OF CALIBRATION. THE AIL PCB WAS RECALIBRATED. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
Description of Event or Problem · 1
DURING SERVICE AT BAXTER, A TECHNICIAN DISCOVERED A FAILURE (B)(4) IN THE EVENT HISTORY THAT WAS DUE TO THE AIL PCB (AIR IN LINE PRINTED CIRCUIT BOARD) WAS OUT OF CALIBRATION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION FOR THIS DEVICE IS 5.09.90, CATEGORIZED AS REMEDIATED.