FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL

MDR report key: 1792178 · Received August 10, 2010

Report

Report Number
6000001-2010-02465
Event Type
Malfunction
Date Received
August 10, 2010
Date of Event
July 20, 2010
Report Date
July 20, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ROOT CAUSE FOR THIS ISSUE IS CURRENTLY BEING INVESTIGATED THROUGH CAPA (CORRECTIVE AND PREVENTIVE ACTION) INVESTIGATION MDQ-CAPA-(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE ASSIGNABLE CAUSE WAS THE AIL PCB WAS OUT OF CALIBRATION. THE AIL PCB WAS RECALIBRATED. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

DURING SERVICE AT BAXTER, A TECHNICIAN DISCOVERED A FAILURE (B)(4) IN THE EVENT HISTORY THAT WAS DUE TO THE AIL PCB (AIR IN LINE PRINTED CIRCUIT BOARD) WAS OUT OF CALIBRATION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION FOR THIS DEVICE IS 5.09.90, CATEGORIZED AS REMEDIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1