COLLEAGUE CXE VOLUMETRICINFUSION PUMP
Report
- Report Number
- 6000001-2010-02464
- Event Type
- Malfunction
- Date Received
- August 10, 2010
- Date of Event
- July 8, 2010
- Report Date
- July 9, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Removal / Correction Number
- 6000001-3/15/05-007-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4).THE AIL PCB HAS BEEN RECALIBRATED.A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). THE ROOT CAUSE FOR THIS ISSUE IS CURRENTLY BEING INVESTIGATED THROUGH CAPA (CORRECTIVE AND PREVENTIVE ACTION) INVESTIGATION MDQ-CAPA-(B)(4).
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 810:11 THAT SHOWED UP FOUR TIMES DURING ONE TESTING. THE REPORTED CONDITION WAS IDENTIFIED DURING BIOMEDICAL SERVICE. DURING SERVICE AT BAXTER, IT WAS DISCOVERED FAILURE CODE 810:11 WAS CAUSED BY THE AIL PCB (AIR IN LINE PRINTED CIRCUIT BOARD) OUT OF CALIBRATION. THERE WAS NO DOCUMENTED PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION FOR THIS DEVICE IS 6.13.90, CATEGORIZED AS REMEDIATED.