FDA Adverse Event Malfunction Summary report: N

TAG 3.7MM ROD II

MDR report key: 1792156 · Received August 10, 2010

Report

Report Number
1219602-2010-00206
Event Type
Malfunction
Date Received
August 10, 2010
Date of Event
July 2, 2010
Report Date
July 7, 2010
Manufacturer
SMITH & NEPHEW ENDOSCOPY MANSFIELD MANUFACTURING
Product Code
MBI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A SMALL PIECE OF THE ANCHOR WAS RETURNED FOR EVALUATION. THE ANCHOR BROKE AT THE SUTURE EYELET. THE ANCHOR APPEARS TO BE SLIGHTLY BENT. THE BREAK IS ANGULAR IN NATURE. THERE IS NO DAMAGE TO THE PROXIMAL END OF THE ANCHOR. CONDITION OF THE ANCHOR IS CONSISTENT WITH OFF AXIS INSERTION. (B)(4).

Description of Event or Problem · 1

DURING SHOULDER SURGERY, DR WAS INSERTING A TAG SUTURE ANCHOR USING ARTHREX TIGER WIRE AS THE SUTURE MATERIAL. AS HE WAS PULLING ON THE SUTURE THE TAG ANCHOR FRACTURED IN HALF AND PART WAS LEFT IN THE PATIENT. A BACK-UP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAG 3.7MM ROD II TAG 3.7MM ROD II MBI SMITH & NEPHEW ENDOSCOPY MANSFIELD MANUFACTURING 014349 50319366

Patients

Seq Age Sex Outcome Treatment
1