AUTOMATED PD SET W/CASSETTE4 PRONG
Report
- Report Number
- 1423500-2010-02349
- Event Type
- Malfunction
- Date Received
- August 10, 2010
- Date of Event
- July 6, 2010
- Report Date
- July 19, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SAMPLE LOT NUMBER IS UNKNOWN AT THIS TIME. THE SAMPLE AVAILABILITY IS UNKNOWN AT THIS TIME. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS OBTAINED.
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. HAVE BEEN COMPLETED TO REFLECT SAMPLE AVAILABILITY. THE CAUSE WAS UNDETERMINED. THE BATCH REVIEW WAS NOT PERFORMED BECAUSE THE LOT NUMBER IS UNKNOWN. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
DURING EVALUATION OF THE HOMECHOICE (HC) DEVICE A SYSTEM ERROR (SE)2240 ALARM WAS IDENTIFIED IN THE DEVICE LOGS WHICH OCCURED ON (B)(6) 2010. A SE 2240 IS A NON-RECOVERABLE ALARM THAT OCCURS WHEN THE DEVICE HAS DETECTED AIR BEING DRAWN CONTINUOUSLY INTO THE DISPOSABLE. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOMATED PD SET W/CASSETTE4 PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |