FDA Adverse Event Malfunction Summary report: N

HOMECHOICE CYCLER-REFURBISHED

MDR report key: 1792036 · Received August 10, 2010

Report

Report Number
1423500-2010-02351
Event Type
Malfunction
Date Received
August 10, 2010
Date of Event
June 27, 2010
Report Date
July 19, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MEDWATCH WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE PASSED THE HOMECHOICE RETURN INSTRUMENT TEST / EVALUATION (RITE) ELECTRICAL TEST AND WAS FOUND TO MEET FUNCTIONAL SPECIFICATIONS RELATIVE TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED VIA THE DEVICE LOGS. THE CAUSE OF THE IIPV WAS DETERMINED TO BE: INSUFFICIENT DRAIN-ONE OR MORE CYCLE(S) ADVANCES TO NEXT FILL WHEN SLOW/ NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). ADDITIONAL INFORMATION: PRODUCT SURVEILLANCE FOLLOWED UP WITH THE PERITONEAL DIALYSIS (PD) NURSE AND PROVIDED RESULTS OF DEVICE EVALUATION. NO OVERFILL SYMPTOMS WERE REPORTED BY HP PER NURSE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. HP CONTINUES PD THERAPY USING THE CYCLER.

Description of Event or Problem · 1

TWO INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATIONS WERE IDENTIFIED IN THE THERAPY LOG OF A HOMECHOICE (HC) DEVICE. THIS IS REPORT 1 OF 2. THE IIPV OCCURRED ON (B)(6) 2010 DURING DRAIN CYCLE 5. THE PROGRAMMED FILL VOLUME WAS 1800ML AND THE TOTAL DRAIN VOLUME WAS 3316ML. THIS DRAIN VOLUME MEETS BAXTER'S IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE CYCLER-REFURBISHED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 73 YR