FDA Adverse Event Malfunction Summary report: N

CAPSURE SENSE

MDR report key: 1791871 · Received August 10, 2010

Report

Report Number
2182208-2010-00432
Event Type
Malfunction
Date Received
August 10, 2010
Date of Event
March 10, 2010
Manufacturer
MEDTRONIC, INC.
Product Code
DTB
PMA / PMN Number
P830061/S034
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY (B)(4) THE FULL LEAD RETURNED AND ANALYZED. PRIMARY ANALYSIS FINDING: NO ANOMALIES FOUND. LEAD MODEL 4574: THE MANUFACTURING DATE CANNOT BE CONFIRMED WITHOUT THE DEVICE SERIAL NUMBER.

Description of Event or Problem · 1

THE REPORT INDICATED DURING THE IMPLANT PROCEDURE, LEAD IMPEDANCE WAS HIGHER THAN 2000 OHMS AND WAS ABNORMAL AFTER SEVERAL TRIES. THEREFORE, THE LEAD WAS WITHDRAWN. A SECOND LEAD WAS UNSUCCESSFULLY ATTEMPTED, AS THE SAME PROBLEM OCCURRED, AND CONSEQUENTLY, THIS LEAD WAS WITHDRAWN AS WELL. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SENSE IMPLANTABLE PACING LEAD DTB MEDTRONIC, INC. 4574 ASKU

Patients

Seq Age Sex Outcome Treatment
1 88 Required Intervention