CONTOUR CURVED CUTTER STAPLER
Report
- Report Number
- 3005075853-2010-04467
- Event Type
- Injury
- Date Received
- August 5, 2010
- Date of Event
- July 7, 2010
- Report Date
- July 14, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- GDW
- PMA / PMN Number
- K040038
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITIONS AND WITH A CARTRIDGE RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED VOID OF STAPLES, WITH THE WASHER COMPLETELY CUT, WITH THE DRIVERS EXPOSED AND WITH THE KNIFE RECESS BELOW THE CARTRIDGE DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH AN ENGINEERING SAMPLE RELOAD AND THE DEVICE FIRED, FORMING ALL STAPLES AND CUTTING AS INTENDED. THE CUT LINE WAS COMPLETE, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE LOCKOUT AND THE RELOAD LOCKOUT WERE FUNCTIONAL. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
.
IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION, ON THE DISTAL TRANSECTION, THE DISTAL STAPLE LINE WAS OPENED. WHEN THE SURGEON WAS CHECKING THE DISTAL STUMP, HE DID NOT SEE ANY STAPLES. SUTURES WERE USED TO COMPLETE THE CASE. THE PATIENT WAS TAKEN BACK INTO THE OPERATING ROOM THAT NIGHT FOR BLEEDING AT THE DISTAL AREA WHERE THE SURGEON HAD USED SUTURES. THERE WAS NO BLOOD TRANSFUSION GIVEN. THE PATIENT IS FINE AND MAY BE RELEASED SOMETIME DURING THE WEEK OF (B)(6) 2010. ADDITIONAL INFORMATION REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR CURVED CUTTER STAPLER | GDW STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC | NA | G4T2OE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other| R | CARTRIDGE |