CONTOUR CURVED CUTTER STAPLER

MDR report key: 1791831 · Received August 5, 2010

Report

Report Number
3005075853-2010-04467
Event Type
Injury
Date Received
August 5, 2010
Date of Event
July 7, 2010
Report Date
July 14, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
GDW
PMA / PMN Number
K040038
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITIONS AND WITH A CARTRIDGE RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED VOID OF STAPLES, WITH THE WASHER COMPLETELY CUT, WITH THE DRIVERS EXPOSED AND WITH THE KNIFE RECESS BELOW THE CARTRIDGE DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH AN ENGINEERING SAMPLE RELOAD AND THE DEVICE FIRED, FORMING ALL STAPLES AND CUTTING AS INTENDED. THE CUT LINE WAS COMPLETE, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE LOCKOUT AND THE RELOAD LOCKOUT WERE FUNCTIONAL. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION, ON THE DISTAL TRANSECTION, THE DISTAL STAPLE LINE WAS OPENED. WHEN THE SURGEON WAS CHECKING THE DISTAL STUMP, HE DID NOT SEE ANY STAPLES. SUTURES WERE USED TO COMPLETE THE CASE. THE PATIENT WAS TAKEN BACK INTO THE OPERATING ROOM THAT NIGHT FOR BLEEDING AT THE DISTAL AREA WHERE THE SURGEON HAD USED SUTURES. THERE WAS NO BLOOD TRANSFUSION GIVEN. THE PATIENT IS FINE AND MAY BE RELEASED SOMETIME DURING THE WEEK OF (B)(6) 2010. ADDITIONAL INFORMATION REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR CURVED CUTTER STAPLER GDW STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC NA G4T2OE

Patients

Seq Age Sex Outcome Treatment
1 Other| R CARTRIDGE