FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ TIP SYRINGE

MDR report key: 17916859 · Received October 11, 2023

Report

Report Number
1911916-2023-00764
Event Type
Malfunction
Date Received
October 11, 2023
Date of Event
September 25, 2023
Report Date
October 4, 2023
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
30382903028321
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY. IT WAS REPORTED THE SOLUTION LEAKS BETWEEN THE BODY AND THE PLUNGER. TO AID IN THE INVESTIGATION, ONE HUNDRED SEVEN SAMPLES IN SEALED PACKAGING BLISTERS WERE RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED, AND NO DEFECTS OR IMPERFECTIONS WERE OBSERVED. TWENTY-FIVE SAMPLES WERE RANDOMLY SELECTED TO TEST FOR LEAKAGE. EACH OF THE TWENT-FIVE SAMPLES PASSED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 302832, LOT 2243878. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED AND A PROBABLE ROOT CAUSE COULD NOT BE DETERMINED. H3 OTHER TEXT: SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD LUER-LOK¿ TIP SYRINGE LEAKAGE OCCURRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DRUG LEAKS BETWEEN THE BODY AND THE PLUNGER. THE DRUG USED IN PREPARATION IS NORMAL SALINE. PLEASE INVESTIGATE WITH SAMPLES OF THE SAME LOT NUMBER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1285457 BD LUER-LOK¿ TIP SYRINGE PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 2243878 30382903028321

Patients

Seq Age Sex Outcome Treatment
1 Unknown