FDA Adverse Event Malfunction Summary report: N

LOANER CORE IMPACTION DRILL

MDR report key: 1791642 · Received July 29, 2010

Report

Report Number
1811755-2010-00858
Event Type
Malfunction
Date Received
July 29, 2010
Date of Event
March 29, 2010
Report Date
March 29, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE WAS RETURNED TO THE MANUFACTURER FOR SERVICE AND EVALUATION. A CONDITION OF THE DEVICE OVERHEATING WAS REPORTED. BASED ON THE REPAIR DETAILS, A POSSIBLE CAUSE FOR THE OVERHEATING WAS A FAILED ROTOR WITH A WORN COUPLER. THE ROTOR, BEARINGS, NOSE CONE, AND BEARING STOP WERE EACH REPLACED. SERVICE REPAIRED THE HANDPIECE AND DID A CLEAN, LUBE, AND ADJUST TO THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDPIECE OVERHEATED DURING AN ORAL SURGERY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER DEVICE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LOANER CORE IMPACTION DRILL SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK