HARMONIC CURVED SHEARS INSERT ACCESSORY
Report
- Report Number
- 2955842-2010-00346
- Event Type
- Malfunction
- Date Received
- July 29, 2010
- Date of Event
- June 29, 2010
- Report Date
- June 29, 2010
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- NAY
- PMA / PMN Number
- K042855
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NURSE
Narratives
THE INSERT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE CLAMP ARM WAS BROKEN OFF AND ONE SIDE OF THE HINGE FEATURE THAT ACCEPTS THE CLAMP ARM PIN IS BENT. THE EVIDENCE IS NOT CONCLUSIVE, HOWEVER, THE DAMAGE IS CONSISTENT WITH HYPEREXTENSION OF THE CLAMP ARM. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NO USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.
IT WAS REPORTED THAT DURING A DA VINCI S ASSISTED CYSTECTOMY PROCEDURE, THE TIP OF THE HARMONIC CURVED SHEARS INSERT ACCESSORY BROKE OFF AND FELL INTO THE PATIENT. THE LOCATION OF THE BROKEN PIECE WAS IDENTIFIED VIA X-RAY AND THE BROKEN PIECE WAS SUCCESSFULLY RETRIEVED DURING THE OPEN SURGICAL PROCEDURE PORTION OF THE PLANNED CASE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC CURVED SHEARS INSERT ACCESSORY | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL, INC. | 400169-07 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | ESU| HCS INSTRUMENT| DA VINCI S SYSTEM & ACCS |