FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1791398 · Received August 6, 2010

Report

Report Number
1720753-2010-02356
Event Type
Malfunction
Date Received
August 6, 2010
Date of Event
May 11, 2010
Report Date
August 6, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE SATA CABLE. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE SYSTEM WOULD NOT SAVE IMAGES. UPON REBOOT, THE SYSTEM WOULD NOT COMPLETE BOOT UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1