FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 1791231 · Received August 2, 2010

Report

Report Number
1824206-2010-09109
Event Type
Malfunction
Date Received
August 2, 2010
Date of Event
July 8, 2010
Report Date
July 8, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE THREE BRAKE CASTERS AND INSTALLED A BRAKE/STEER UPGRADE TO REPAIR THE STRETCHER.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THREE BRAKE CASTERS ROLL SLIGHTLY AND SWIVEL WHEN IN THE BRAKE POSITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILL-ROM, INC. 8000

Patients

Seq Age Sex Outcome Treatment
1