HARMONIC ACE 36CM W ERG HANDLE
Report
- Report Number
- 3005075853-2010-04553
- Event Type
- Malfunction
- Date Received
- August 10, 2010
- Date of Event
- July 14, 2010
- Report Date
- July 20, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). THE ANALYSIS SHOWED THAT THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED. THE DEVICE WAS ACTIVATED WITH THE GENERATOR AND AN ERROR CODE 5 WAS DISPLAYED. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN AN ERROR CODE 5 OR BLADE LOCKOUT LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE.
DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED A DEPLETED BATTERY ALARM, WHICH INTERRUPTED DELIVERY FIVE TIMES ON THE EVENT DATE. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION (B)(4).
IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE WHEN THE DEVICE WAS REMOVED THROUGH THE TROCAR THE SURGEON NOTICED THAT THE ACTIVE BLADE TIP HAD BROKEN OFF OF THE DEVICE AND WAS IN THE PATIENT. THE TIP WAS RETRIEVED WITH LAPAROSCOPIC FORCEPS. THEY COMPLETED THE PROCEDURE WITH A NEW LIKE DEVICE. THERE WAS NO PATIENT CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 36CM W ERG HANDLE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | F4PM8T |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |