FDA Adverse Event Malfunction Summary report: N

EON IMPLANTABLE PULSE GENERATOR

MDR report key: 1790905 · Received August 6, 2010

Report

Report Number
1627487-2010-01572
Event Type
Malfunction
Date Received
August 6, 2010
Date of Event
February 21, 2008
Report Date
February 26, 2008
Manufacturer
ADVANCED NEUROMODULATION SYSTEMS
Product Code
LGW
PMA / PMN Number
P010032
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. IPG PASSES ALL FUNCTIONAL TESTING INCLUDING LEAD RETENTION TESTING. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS.(B)(4). ANS HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

THE PATIENT RECEIVED HER SCS SYSTEM INCLUDING AN IPG AND LEADS ON (B)(6) 2007. IT WAS REPORTED THAT THE LEADS PULLED OUT OF THE IPG. THE PHYSICIAN EXPLANTED AND REPLACED THE IPG ON (B)(6) 2008. THE EXPLANTED IPG WAS RETURNED TO ANS FOR ANALYSIS. NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EON IMPLANTABLE PULSE GENERATOR TOTALLY IMPLANTABLE PULSE GENERATOR LGW ADVANCED NEUROMODULATION SYSTEMS 3716 85246

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention