OCTRODE LEAD KIT, 30CM LENGTH
Report
- Report Number
- 1627487-2010-01333
- Event Type
- Malfunction
- Date Received
- August 6, 2010
- Date of Event
- November 5, 2008
- Report Date
- November 26, 2008
- Manufacturer
- ADVANCED NEUROMODULATION SYSTEMS
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVAL - THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS : THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECS AND NO ANOMALIES WERE FOUND. FAILED CONTINUITY TEST. LEAD WAS KINKED WITH ALL BROKEN WIRES. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. THIS MDR IS BEING SUBMITTED PAST THE 30 DAY REPORTING REQUIREMENT AS PART OF A RETROSPECTIVE REVIEW INITIATED IN RESPONSE TO AN FDA INSPECTION. A RETROSPECTIVE REVIEW OF THE COMPLAINT RECORD DETERMINED THAT ANS MISINTERPRETED THE MDR REGULATIONS IN THIS INSTANCE. ANS HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
THE PT RECEIVED HIS SCS SYSTEM INCLUDING PERCUTANEOUS LEADS ON (B)(6)2007. IT WAS REPORTED THAT THE PT EXPERIENCED A LOSS STIMULATION. LEAD IMPEDANCES WERE MEASURED AND FOUND TO BE HIGH. A FRACTURE ON THE LEAD WAS VISIBLE. THE LEAD WAS REPLACED ON (B)(6)2008. FOLLOW UP ON THE PT FOUND THAT NO FURTHER ISSUES HAVE BEEN REPORTED. THE EXPLANTED LEAD WAS RETURNED TO ANS FOR EVAL. NO FURTHER INFO IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OCTRODE LEAD KIT, 30CM LENGTH | SPINAL CORD STIMULATION LEAD | LGW | ADVANCED NEUROMODULATION SYSTEMS | 3183 | 64147 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |