FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 17908557 · Received October 10, 2023

Report

Report Number
1221359-2023-01537
Event Type
Malfunction
Date Received
October 10, 2023
Date of Event
September 30, 2023
Report Date
October 23, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

D4 UDI: (B)(4) TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 225415 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 225415, TEST BASE PART NUMBER 195-430H / LOT 220852. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 225415 SHOWED THAT THE COMPLAINT RATE IS (B)(4) ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE. H3 OTHER TEXT : SINGLE-USE, DEVICE DISCARDED.

Additional Manufacturer Narrative · 0

D4 UDI: (B)(4). THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION OR UPON RECEIPT OF NEW INFORMATION. H3 OTHER TEXT : SINGLE-USE, DEVICE DISCARDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF TEST PERFORMED ON (B)(6)2023. REPEAT TESTING WAS PERFORMED WHICH GENERATED A NEGATIVE RESULT. ADDITIONAL TESTING WAS PERFORMED USING UNKNOWN BRAND TEST WHICH GENERATED A NEGATIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF TEST PERFORMED ON (B)(6) 2023. REPEAT TESTING WAS PERFORMED WHICH GENERATED A NEGATIVE RESULT. ADDITIONAL TESTING WAS PERFORMED USING UNKNOWN BRAND TEST WHICH GENERATED A NEGATIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331727 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 225415 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 Unknown