EON 16-CHANNEL IPG
Report
- Report Number
- 1627487-2010-01641
- Event Type
- Malfunction
- Date Received
- August 6, 2010
- Date of Event
- August 26, 2009
- Report Date
- August 26, 2009
- Manufacturer
- ADVANCED NEUROMODULATION SYSTEMS
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. (B)(4). ANS HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
THE PATIENT RECEIVED HER SCS SYSTEM ON (B)(6) 2007. IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING LONG RECHARGE TIMES, LOW PARAMETERS, POSITIONAL STIMULATION, AND THE IPG TURNING ON AND OFF. THE SYSTEM WAS REPLACED ON (B)(6) 2009. FOLLOW UP ON THE PATIENT FOUND THAT NO FURTHER ISSUES HAVE BEEN REPORTED. THE EXPLANTED IPG WAS RETURNED TO ANS FOR EVALUATION. NO FURTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EON 16-CHANNEL IPG | TOTALLY IMPLANTABLE PULSE GENERATOR | LGW | ADVANCED NEUROMODULATION SYSTEMS | 3716 | 65829 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |