PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE(SMC) SYSTEM
Report
- Report Number
- 2953144-2010-01539
- Event Type
- Injury
- Date Received
- August 4, 2010
- Date of Event
- June 29, 2010
- Report Date
- July 12, 2010
- Manufacturer
- ABBOTT VASCULAR REDWOOD CITY
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). DEVICE #2: PART #12673003, LOT #90005-6H INDICATED IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS PROVIDED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.
DEVICE ISSUE #1: NEEDLE-TO-CUFF MISS. TIME OF DEVICE ISSUE: DURING VESSEL CLOSURE. ADVERSE EVENT: FAILURE TO ACHIEVE HEMOSTASIS. IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, WHEN THE NEEDLE PLUNGER WAS REMOVED, A CUFF MISS OCCURRED. THE PROGLIDE WAS REMOVED AND A SECOND PROGLIDE DEVICE WAS USED WITH THE SAME RESULTS. HEMOSTASIS WAS ACHIEVED USING MANUAL COMPRESSION. THERE WERE NO REPORTED PATIENT ADVERSE EFFECTS. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE(SMC) SYSTEM | MGB | ABBOTT VASCULAR REDWOOD CITY | NA | 90005-6H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | DEVICE #2, PROGLIDE, PART #12673-03, LOT #90005-6H |