DUAL 4 LEAD EXTENSION, 60 CM
Report
- Report Number
- 1627487-2010-01551
- Event Type
- Injury
- Date Received
- August 6, 2010
- Date of Event
- December 28, 2007
- Report Date
- January 3, 2008
- Manufacturer
- ADVANCED NEUROMODULATION SYSTEMS
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE 2 OF 2. THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. LEAD SEGMENT IS KINKED AT THE STRAIN RELIEF. ONE WIRE WAS BROKEN AND EXTENSION FAILS CONTINUITY TESTING. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. THIS MDR IS BEING SUBMITTED PAST THE 30 DAY REPORTING REQUIREMENT AS PART OF A RETROSPECTIVE REVIEW INITIATED IN RESPONSE TO AN FDA INSPECTION. A RETROSPECTIVE REVIEW OF THE COMPLAINT RECORD DETERMINED THAT ANS MISINTERPRETED THE MDR REGULATIONS IN THIS INSTANCE. ANS HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REFERENCE MANUFACTURER REPORT: 1627487-2010-01538.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DUAL 4 LEAD EXTENSION, 60 CM | SPINAL CORD STIMULATION EXTENSION | LGW | ADVANCED NEUROMODULATION SYSTEMS | 3346 | 57019 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |