FDA Adverse Event Injury Summary report: N

EON MINI IMPLANTABLE PULSE GENERATOR

MDR report key: 1790687 · Received August 6, 2010

Report

Report Number
1627487-2010-01668
Event Type
Injury
Date Received
August 6, 2010
Date of Event
September 6, 2008
Report Date
September 8, 2008
Manufacturer
ADVANCED NEUROMODULATION SYSTEMS
Product Code
LGW
PMA / PMN Number
P010032
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. THIS MDR IS BEING SUBMITTED PAST THE 30 DAY REPORTING REQUIREMENT AS PART OF A RETROSPECTIVE REVIEW INITIATED IN RESPONSE TO AN FDA INSPECTION. A RETROSPECTIVE REVIEW OF THE COMPLAINT RECORD DETERMINED THAT ANS MISINTERPRETED THE MDR REGULATIONS IN THIS INSTANCE. ANS HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

THE PATIENT RECEIVED HIS SCS SYSTEM ON (B)(6) 2008. ON (B)(6) 2008, IT WAS REPORTED THAT THE PATIENT LOST STIMULATION WHEN HE REMOVED THE PROGRAMMER WAND FROM THE IPG. THE POWER DISPLAY SHOWED THAT THE IPG WAS FULLY CHARGED. THE LEADS WERE REPOSITIONED BECAUSE IT WAS SUSPECTED THAT THE POWER REQUIREMENTS WERE CAUSING THE IPG TO MALFUNCTION. THE IPG FUNCTIONED PROPERLY AFTER THE LEADS WERE REPOSITIONED. NO DEVICE WAS EXPLANTED AND NO DEVICES WERE RETURNED FOR ANALYSIS. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EON MINI IMPLANTABLE PULSE GENERATOR TOTALLY IMPLANTABLE PULSE GENERATOR LGW ADVANCED NEUROMODULATION SYSTEMS 3788 172207

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention