FDA Adverse Event Injury Summary report: N

OCTRODE PERCUTANEOUS LEAD

MDR report key: 1790657 · Received August 6, 2010

Report

Report Number
1627487-2010-01660
Event Type
Injury
Date Received
August 6, 2010
Date of Event
July 17, 2008
Report Date
July 17, 2008
Manufacturer
ADVANCED NEUROMODULATION SYSTEMS
Product Code
LGW
PMA / PMN Number
P010032
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE 2 OF 2. THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. THE LEAD IS SEVERELY KINKED WITH ALL WIRES BROKEN APPROX 17.5CM FROM THE STIM END. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. THIS MDR IS BEING SUBMITTED PAST THE 30 DAY REPORTING REQUIREMENT AS PART OF A RETROSPECTIVE REVIEW INITIATED IN RESPONSE TO AN FDA INSPECTION. A RETROSPECTIVE REVIEW OF THE COMPLAINT RECORD DETERMINED THAT ANS MISINTERPRETED THE MFR REGULATIONS IN THIS INSTANCE. ANS HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REF MFR'S REPORT: 1627487-2010-01659.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OCTRODE PERCUTANEOUS LEAD SPINAL CORD STIMULATION LEAD LGW ADVANCED NEUROMODULATION SYSTEMS 3186 68280

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention